PAPER / ARXIV:2609.18589
Natascha Niessen , Ana Beatriz Solana , Carolin M. Pirkl , Tim Sprenger , Hannah Eichhorn , Veronika Spieker , Wenqi Huang , Rolf F. Schulte , Florian Wiesinger , Tobias C. Wood , Marion I. Menzel , Julia A. Schnabel on behalf of the PREDICTOM consortium
RESUMO
MPnRAGE enables multiple inversion contrast images in a single scan, allowing quantitative T1 mapping, tissue nulled contrasts, and standard MPRAGE synthesis. However, current 3D scan times remain clinically impractical, motivating accelerated 3D MPnRAGE. This work provides a highly accelerated Cartesian 3D MPnRAGE sequence with joint implicit neural representation (INR) reconstruction. The sequence uses a tailored view-ordering strategy, flip angle schedule and complementary variable-density Poisson-disk undersampling. Calibration data acquired during otherwise unused delay time are used for sensitivity map estimation and complementary high-frequency sampling. Ten INR-reconstructed inversion images at 1.5 mm$^3$ are evaluated against fully sampled references via retrospective undersampling. Prospectively accelerated 1 mm$^3$ images at R = 20 (5.39 min) demonstrate clinical feasibility. INR reconstruction outperforms subspace and iterative local low rank reconstruction on highly accelerated data. The proposed highly undersampled 3D Cartesian MPnRAGE with INR reconstruction generates multiple high-quality inversion contrasts in substantially reduced scan time. Scan-specific INR reconstruction improves image quality while reducing reconstruction time versus state-of-the-art methods.
NO MESMO MAPA