PAPER / ARXIV:2609.06230
T. W. Clark , B. Sárközi , A. Dombi , Á. Kurkó , D. Nagy , A. Simon , D. Varga , P. Domokos , A. Vukics
RESUMO
Precisely timed, multi-device control in atomic, molecular and optical physics is provided by specialized real-time systems, which impose their own terms on the description of the experiment: a program over a global clock, in which every stage boundary is an absolute time computed from everything preceding it. Such descriptions are tightly coupled -- changing one stage affects all later ones -- and are correspondingly hard to reuse, inspect, or move elsewhere. We introduce Wigner Time, a Python package in which the experimental procedure is instead represented as data: a table of timed updates, assembled by composing functions and referred to named points in the experiment rather than to absolute instants. Hardware enters only at a final conversion step, leaving the description readable and back-end agnostic. Wigner Time has run two cold-atom setups for more than two years, where replacing a hand-written real-time program improved the achievable temporal resolution five-fold; the design applies to any domain requiring precise multi-device timing.